Software for clinics & diagnostic labs. From ₹399/mo, 10-day free trial. See plans →

Internal Quality Control (IQC) Record and Levey–Jennings Chart

Required by: ISO 15189:2022 Cl. 7.3.7.2 — Internal quality control

Daily proof that each analytical system was in control on the day the patient result was produced.

Checked against NABL 112A (Issue 01, 18-Dec-2024) · page updated 23 September 2026

Layout of a compliant internal quality control (iqc) record and levey–jennings chart for an Indian medical laboratory, showing every field required under ISO 15189:2022 Cl. 7.3.7.2 — Internal quality control
What a compliant internal quality control (iqc) record and levey–jennings chart looks like. Every field shown is required — an assessor reads the record, not the intention behind it.

How often it must be recorded

Two levels of QC on the day of testing, then one level every shift / 8 hours, run at fixed times during operational hours.

Every field this record must carry

An assessor reads the record, not the intention behind it. A field left off the form is a finding, however well the work was actually done.

  • Analyte and instrument
  • QC lot and level
  • Date, shift and fixed run time
  • Observed value
  • Laboratory's own mean and SD
  • LJ plot point
  • Accept / reject decision against defined criteria
  • Outlier, trend and shift analysis
  • Corrective action taken
  • Monthly mean, SD and %CV

What NABL 112A actually says

NABL 112A: two levels of QC shall be included on the day of performing the test and subsequently one level every shift/8 hour; controls shall be run at fixed timings during operational hours. The laboratory shall derive its OWN mean and SD using a minimum of 20 data points to plot an LJ chart, shall review LJ charts daily, and shall calculate monthly mean, SD and %CV. Where a control has a short shelf life (the document gives CBC as the example) the manufacturer's assigned mean and SD may be used, but the laboratory must still calculate its own imprecision as %CV and keep it within the acceptable range. A minimum of six months of IQC data is recommended to calculate routine imprecision, updated annually where possible.

Source: NABL 112A — Specific Criteria for Accreditation of Medical Laboratories, Issue 01, 18-Dec-2024, 108pp. Downloaded from nabl-india.org 2026-09-23. Paraphrased with attribution. ISO 15189:2022 clause numbers and titles are cited as references; the standard's text is copyrighted and is not reproduced here.

What an assessor actually asks for

Reasoning from the clause, not from a published checklist — NABL does not publish assessor scripts. Treat this as how the requirement is normally evidenced.

  1. The record itself, for a period they choose — not a blank template.
  2. Completeness across the period. Gaps are the finding; a perfect January and an empty March reads worse than a consistently modest record.
  3. Who recorded and who verified, with dates. An unsigned record is not evidence that the work was checked.
  4. The linked procedure — the SOP that says this record must exist, at this frequency, in this format.
  5. What happened when it went wrong. Every record above has a failure path, and the action taken is usually the part that is missing.

Records this one depends on

Where the software fits

HealthFlow is lab software, not an accreditation body. It is not NABL accredited and using it does not make a laboratory accredited or compliant — the records above are the laboratory's own responsibility.

What it does do is produce some of this evidence as a by-product of normal work: turnaround time is measured per test on the owner dashboard, reports are versioned and locked on authorisation with dual signatures, every report access is written to an append-only audit trail enforced at the database, retained two years, and all of it exports at any time.

Ask what it records for you See all 22 records →

WhatsApp us Call
Book a demo